Infant Incubator | #Baby Care 5A
Brief Introduction
Temperature Control Modes
●Tem. control mode:Skin & air temperature control mode
●Air mode control temp. range:20° C-37° C
●Air mode control override temp. range:37° C-39° C
●Air mode control accuracy:≤±1.0°C
●Skin mode control temp. range:34° C-37° C
●Skin mode control override temp. range:37° C-38° C
●Skin temp. accuracy: ±0.3°C
�?/span>With Dual-skin temp. monitoring
Performance
●Air flow velocity across mattress:< 10 cm/sec
●Temp. rise time at 22 °C (72 °F) ambient:<20min (From 22 °C�?0% RH,to 35 °C)
●Temp. variability:< 0.3°C
●Temp. overshoot:< 0.5 °C maximum
●Temp. uniformity with a level mattress:< 0.5°C
●Operating noise level in hood:≤45dBa
●Carbon Dioxide (CO2) Level:<0.5%
Servo Humidity Option
●Humidity control operating time without refilling:24hours
●Humidity control range:30%-95% RH
●Humidity display accuracy:±5%RH
●Humidity control accuracy:±5% RH
●Humidity control reservoir capacity:1500ml
Servo Oxygen Option
●Oxygen control accuracy of full scale:± 2%
●Oxygen control range:21 % to 65%
●Oxygen display accuracy (21% calibration):± 5%
●Oxygen display accuracy (100% calibration):± 3%
●Oxygen display resolution:1%
Scale Option
●Weight range:300g to 8000g
●Weight display resolution:1 g
●Weight accuracy:±10 g
Operating Environment
●Temperature:20° to 30° C
●Humidity:10 to 95% RH
●Air Velocity:Upto 0.3 m/sec
Storage/Shipping Information
●Temperature: -26° to 60° C
●Humidity:0 to 96% RH