Brief introduction:
servo-controlled by micro-computer; double wall hood an reduce the heating losing and the effect of outer heat resource; automatic curtain can reduce the heat losing ; inclination of infant bed can be adjusted; quite operation windows can reduce the disturbance to the baby during nursing; touch switch panel with LED digital display, set-temp, real-temp and auxiliary temperature are displayed separately; >37ºC temperature setting; multiply self-checking alarms with audible and visible; heating power rate indicates 0-100%; whole aluminum water tank, and it can be filled with water or discharge; RS 232 connector, data output.
Standard configure:
Hood, infant bed, temperature controller, skin temperature sensor, water tank, air filter, Guardrail, I.V. Pole, cabinet.
Technical specification:
Power supply: AC220V,50HZ
Power input: 600VA
Air Temperature control range:25ºC~37ºC can override to 39ºC
Skin Temperature control range:34ºC~37ºC can override to 38ºC
Sensor precision: <=0.3ºC
Changing of temperature: <=0.5ºC
Uniformity of temperature: <=0.8ºC
Temperature rising time: <=30min
Noise level: <=55dB(A)
Inclination of bassinet: 0º~5º
Alarming:
audible and visible alarm indication, and it can indicate the real temperature and cut off the power supply of heater (except for the low deviation alarm), meanwhile, the skin temperature indication window can indicate the alarm code. For the alarm 1-6, the relevant light is on, and for the alarm 7, all alarming lights are on; all alarm sound can be cancelled except for the power failure alarm and the system alarm.
Over-temp alarm: when the air temperature control <=38ºC(setting temperature is less than 37ºC ), or<=40ºC(setting temperature is more than 37ºC);skin temperature control <=40ºC;
Upper Deviation alarm: air temperature control +3.0ºC, skin temperature control +1.0ºC;
Lower Deviation alarm: air temperature control -3.0ºC, skin temperature control -1.0ºC;
Sensor alarm: when the sensor is disconnected, open circuit, short circuit or put on the wrong place;
Fan alarm: when the fan is blocked, speed <=1000rpm or stop working;
Power failure alarm: when there is no power or the power cord is disconnected.
System alarm: when there is E2ROM malfunction inside of temperature controller.
Package:
Gross Weight:Main body 66Kg / Cabinet 68Kg
Measurement:Main body 1085mmx640mmx900mm
Cabinet:1125mmx665mmx715mm |